Rice Lake Digital/Diagnostic - iQube User Manual

Page 50

Advertising
background image

Appendix

47

Virtui

_____________

4. What communication method is being used?

RS-232___________

RS-485___________

Fiber–optic_________

R/F______________

5. What is the

iQUBE

configuration?

6. Do you have the manual for all the equipment in use?

7. After connection, is there Voltage at the

iQUBE

?

8. Is D1 On or Off on the connector board?

9. If using a

920i

, continue to step 10.

If using

Revolution III

, access cell monitoring Via the interactive tab and proceed to step 18

If using

Virtui

, access cell monitoring and proceed to step 18

10. What is the distance from the

920i

to the scale?

11. What port is the

iQUBE

on?

If using port 3, ensure printer ports are not still set to the default port 3.

12. With the communications on, using

TESTCOM

, are D5 and D6 or D7 and D8 flashing?

13. If using

iRev

, is the

CONFIG

already set up and ready to download?

14. Set up the diagnostic key.

15. Does the download complete to the

920i

without error?

16. Did the

920i

send the data to the

iQUBE

?

17. Does the Cal-Match work?

18. When the scale is empty, are all the current readings near the captured zero?

If yes, go to step 19.

If no, go to step 20.

19. With weight on the scale, do all of the current readings rise to the same degree?

If yes, go to step 20.

If no, go to step 21.