CEL-63x Environmental & Occupational Noise Meter

53

Casella CEL

Other functions listed under L

eq

are as follows:

L

AIeq

is an integrating A-weighted measurement of impulse L

eq

.

Integrated measurements provide an L

eq

measurement that assesses

exposure to noise levels, and are used in the calculation of the personal
noise exposure L

EP

as defined by ISO 1999.

L

C

– L

A

is a measurement that is used in the HML method, where

separate values for high, medium and low frequencies provide a better
estimate.

L

AeqT80

is an A-weighted measurement of L

eq

that has a threshold setting

of 80 dB. This is used for ACGIH in the USA only.

Maximum and Minimum SPLs

This screen allows you to select among the available maximum
and minimum SPL functions.

The subscripted letters identify the frequency weighting and the
time constant to be used for the measurement, and whether the
measurement is a maximum or a minimum.

Peaks, L

Avg

and other functions

Peaks

The Peaks functions are L

APk

, L

CPk

and L

ZPk

.

These functions record the peak levels of noise using A, C and Z
weighting.

L

Avg

L

Avg

is the time-averaged noise level with an applied exchange rate, measured

during the run. The two options are to use exchange rates Q5 (5 dB) or Q4
(4 dB).

Other functions

L

AE

is the A-weighted exposure level (previously known as SEL). It is the

sound pressure level that would contain the same amount of energy in
one second that the actual noise has during the whole measurement
period.

L

Tm3

and L

Tm5

functions are Taktmaximal A-weighted fast or impulse

measurements.

Taktmaximal measurements are specified in German noise standards
(the DIN standard). They integrate noise over a 3 second or a 5 second