CEL-63x Environmental & Occupational Noise Meter

61

Casella CEL

6.10 Time weightings

Fast (F), Slow (S) and Impulse (I) according to IEC 61672-1: 2002.

6.11 Correction filters

Built-in correction filter for random incidence sound pressure fields.

6.12 Reference direction

For free-field measurements the reference direction is perpendicular to the
microphone diaphragm.

6.13 Reference conditions

23 °C air temperature

50 % relative humidity (RH)

101.325 kPa atmospheric pressure

Nominal reference level = 114.0 dB at 1 kHz

6.14 Operating environmental conditions

Humidity

5 % to 90 % RH non-condensing

Temperature

–10 °C to +50 °C (Class 1)

0 °C to +40 °C (Class 2)

Pressure

65 kPa to 108 kPa

6.15 Effects of temperature

Electrical stability of the instrument <±0.2 dB over the range –10 °C to +50 °C

6.16 Effects of humidity

Less than ±0.5 dB over the range 25 % to 90 % RH (non-condensing) relative to
the value under reference conditions.

6.17 Storage environmental conditions

Humidity

0 % to 90 % RH non-condensing

Temperature

–20 °C to +60 °C

Pressure

65 kPa to 108 kPa