29 available data sets – Casella CEL CEL-63X User Manual

Page 64

Advertising
background image

CEL-63x Environmental & Occupational Noise Meter

64

Casella CEL

impedance should be as high as possible, and a coaxial
connection cable of between 0.5 metres and 10 metres length
should be used.

The AC output corresponds to the Z-weighted response of the
instrument.

DC output

2.5 mm stereo jack (barrel ground, ring DC output) provided
for chart recorders, loggers and so on. Some offset and scaling
is necessary on the measuring system for accurate
measurements.

Approximately 1.4 V DC full-scale output corresponding to
140 dB. Output impedance is approximately 2.2 kΩ. Load
impedance should be as high as possible.

The DC output corresponds to the A-weighted, fast time
response of the instrument.

Note that the DC power input ground must be kept isolated
from any signal ground.

6.29 Available data sets

A) Broadband Cumulative and Periodic Results

For each complete measurement run the instrument stores the following:

Run start Date and Time

Run duration

Run overload time

Run pause times

Run Id

Instrument serial number

Current measurement setup data

The Last calibration before run started results

The First calibration after end of run results

The Overload and Battery fail flags

The Environmental indices LDN, LDEN, CNEL

In periodic mode the following set of results is produced for each of the period
intervals during a run.

In cumulative mode the following single set of results is produced covering the
whole run.