3.3.6 Direct Mode Mounting Using Track Mount Hardware

1. Once the installation site selection process described in section 3.2 is complete,

prepare the pipe where the first sensor will be mounted.

2. To prepare the pipe, temporarily position one track mount hardware assembly on each

side of the pipe where you intend to permanently mount them as shown below.

Ensure that pipe surface is smooth without any raised areas (seams, etc.) With a

pencil, marker or chalk, draw a rectangle around each track mount assembly. Remove

the hardware from the pipe and clean and de-grease the area within the rectangles. Use

the small sanding block provided with the installation hardware as necessary to

remove any grit, corrosion, rust, loose paint or other contaminants. Be sure to wipe

the surface clean after sanding. The cleaned surface should extend at least ½” beyond

the length and width of each assembly.

3. Re-install the two track mount hardware assemblies on the pipe and loosely secure

them in place with the mounting straps provided. Align the assemblies such that they

are positioned on top of the cleaned surfaces.

IMPORTANT NOTE

Direct mode mounting requires that transducers be installed on opposite sides of the pipe. For

horizontal pipes, the transducers should be located at the 3 o’clock and 9 o’clock positions.

i

DIRECT MODE SPACING

9

2

0 1

4 5

3

8

6 7

14

11

10

13

12

16

15

18

17

20

19

21

23

22

24

25

REFLECT MODE SPACING

9

REF5

8

6 7

14

12

11

10

13

16

15

18

17

20

19

21

23

22

24

25

O.D. RANGE 1/4"-5"

TRACK 400

TRACK 400
O.D. RANGE 1/4"-5"

Mark pipe around each

track and clean within

the marked area.

Advertising