Using the emery cloth on the steel wire coil, Attaching the steel wire coil to the binding posts – Super Systems CAT-100 User Manual

Page 15

Advertising
background image

CAT-100 Operations Manual

Super Systems Inc.

Page 15 of 51

6. Close the access port valve.

7. Remove the insertion rod completely from the access port.


NOTICE!

Do NOT quench in water; air cool only. If a method of cooling other than air cooling is utilized,

test results will be unreliable.


Using the Emery Cloth on the Steel Wire Coil

Before the wire coil is attached to the CAT-100 for analysis, the two ends of the wire should be

cleaned. An emery cloth is provided for this purpose. It is recommended that each wire end be

lightly cleaned with emery cloth, as pictured in Figure 15.

Figure 15 - Using emery cloth on wire coil


IMPORTANT!

The correct type of emery cloth must be used for acceptable results. SSi sells a fine grade,

150 grit emery cloth for use in this process. See the Replacement Parts list (page 7) for more

information.


Attaching the Steel Wire Coil to the Binding Posts

Attach one of two wire coil leads to the hole on the right side of one of the two binding posts. Be

sure to insert the coil lead until it touches the back of the post (note the contact points in Figure

16). Then tighten the post to secure the lead. Repeat this procedure for the second lead.


IMPORTANT!

Each lead must touch the back of the post to ensure the correct length of insertion. Note the red

arrows in Figure 16.