Login, Calibration, Login calibration – Super Systems CAT-100 User Manual

Page 23: Ee login, E 23), Figure 32)

Advertising
background image

CAT-100 Operations Manual

Super Systems Inc.

Page 23 of 51

Figure 32 - About Screen

LOGIN

The purpose of the LOGIN option (Figure 33) is to allow the user to access Supervisor and

Configuration functions. Enter the Supervisor pass code (default is 1) to log in with the

Supervisor access level. Enter the Configuration pass code (default is 2) to log in with the

Configuration pass code. Functions available with each access level are shown in the CAT-100

Menu Navigation section on page 19.

Figure 33 - LOGIN Option

Calibration

A circuit calibration is required once per year and recommended every four months. The

required annual calibration is performed by SSi, and a certificate is issued for this calibration. A

circuit calibration can be performed by the end-user by using the Calibration Bridge provided by

SSi when the CAT-100 is first shipped.

An important guideline is as follows: A circuit calibration should be performed at any time the

“Error in %” found in this procedure is greater than 0.5 or less than -0.5.

NOTE: Information on ordering a replacement Calibration Bridge can be found in the

Replacement Parts section (page 6).

Advertising