Super Systems Inc.

Page 48 of 51

Problem

Possible Causes

Possible Corrective Actions

% Carbon indicated by CAT-100 differs by more

than 0.03% from Process Variable shown by

instrumentation at the time wire was exposed to

testing atmosphere

(Continued)

Incorrect Coil Factor entered

Ensure that the correct Coil Factor is entered

during the testing procedure; see Entering the

Coil Factor section on page 8.

Carbon Potential is not within the effective range

of 0.1% to 1.3%

Actual Carbon Potential is not within the effective

range of the unit

Soak time too long or too short

% Carbon shown must be within the effective

range in order to be reliable. A different method

of measuring Carbon Potential may be needed in

such a case.

Re-test using an unused wire. Ensure that soak

times shown in Table 2 are applied as indicated.

Wire becomes detached from insertion rod

during soak in testing atmosphere

Unsecured attachment

Before inserting rod into testing atmosphere,

verify that wire coil is in place securely for

testing, that rod screw is securely tightened, and

that protective cage is securely in place on the

rod

Reading is not stored when “Store” is pressed on

CAT-100 touch screen

Pressure on screen not hard enough

Interference from protective film over screen

No reading to store

Internal storage error

Increase pressure of contact with screen. Take

care not to cause damage to screen or protective

film.

Verify that spacing between protective film and

touch screen allows for contact with touch screen

Verify that a test result was generated when wire

coil was connected to binding posts on CAT-100.

If not, reconnect wire to binding posts and

observe reading again.


Call SSi Technical Support at

(513) 772-0060

Test results not displaying correctly in Test

Results screen

Internal storage error

Call SSi Technical Support at

(513) 772-0060