IMPORTANT!

The maximum allowable “Error in %” range is ±05. If the “Error in %” is greater than 0.5 or less

than -0.5 after you have pressed the

button more than twice, contact SSi at (513) 772-0060

for further assistance.



Press the

button to return to the menu list.

Coil TC Calibration

The CAT-100 contains a thermistor used to measure ambient temperature in the area near the

binding posts to which the steel wire coil is attached for carbon analysis. This thermistor must

be calibrated annually.

To calibrate, first press the Coil TC Calibration button. Using a digital thermometer with

accuracy of ±0.1°C (±0.18°F), measure the ambient temperature within ¼” of the thermistor

(see Figure 36). Enter this value in the Coil TC Calibration screen (Figure 37). The “Coil Temp”

value will change to the value you entered.

Figure 36 - Location of external thermistor

Figure 37 – Coil TC Calibration screen

Press the

button to return to the menu list.

Comms Setup (Communications Setup)

The Comms Setup screen allows you to change communications settings of the serial and USB
connections for the CAT-100. Use the

and

buttons to change which setting is selected.

Press the

button to change a setting. In most cases, these settings will not need to be

changed from their defaults.