Testing Procedure

Unpacking and Charging the CAT-100

Carefully remove the packing tape from the box in which the CAT-100 was shipped and remove

the CAT-100, right side up. Using proper lifting technique, place the CAT-100 on a level surface.

If you are going to use the CAT-100 without electrical outlet power, allow the unit to charge

overnight before use. Do this by simply plugging the unit into an AC electrical outlet and

allowing it to charge. Make sure that the unit is charging with the power turned OFF.

Entering the Coil Factor

The Coil Factor for the wire coil is used in calculating Carbon Potential and is based on

variations in coils in a batch. The recommended Coil Factor is printed on each bag of coils and is

the same for all coils in a batch. Before using the coil in the testing process, follow this

procedure for entering and verifying the Coil Factor. It is essential that this procedure be

followed correctly.

First, note the recommended Coil Factor printed on the bag of coils that the coil you are using

came from. If you do not have the bag, assume a Coil Factor of 100.

Next, log in with Configuration Mode

on the CAT-100. This is done by

opening the “LOGIN” screen from the

main menu and entering the

Configuration pass code (which is ‘2’

by default).

Figure 3 – Entering Configuration pass code (default is ‘2’)

Open the Detail menu from the main

menu. The Detail screen will be

shown, and the Coil Factor will be

displayed in a box labeled “Factor”.

Tap on that box.

Figure 4 – Detail screen with Coil Factor box