Watson-Marlow 620UN, 620U, 620SN, 620S User Manual

41

16.5 Remote stop

The 620UN pump can be started and stopped with a remote switch between the 5V
terminal and the i/p terminal of the Run/stop input, using an open=run or
open=stop command sense. It also operates with a logic input between 5V and 24V
on the i/p terminal of the Run/stop input. If remote switching is disabled, neither the
remote run/stop nor direction signal inputs will affect the pump state.

The 620U pump can be started and stopped with a remote switch between pin 7 and
pin 19 using an open=run or open=stop command sense. It also operates with a
logic input between 5V and 24V on pin 7. If remote switching is disabled, neither the
remote run/stop nor direction signal inputs will affect the pump state.

UN, U

U

UN

In the first screen of the Setup menu select Remote stop using the UP and
DOWN keys. Press ENTER to confirm your choice.
The pump displays a screen allowing the user to disable the remote stop fea-
ture. Use the UP and DOWN keys to choose Yes or No and press ENTER to
confirm the decision.
If No is chosen, the pump asks the user to make a further choice, according to
whether the pump is required to run on an open or closed remote switch:
Open=stop or Open=run. Choose using the UP and DOWN keys and press
ENTER to confirm. The pump briefly displays a confirmation screen and returns
the user to the first screen of the Setup menu.
If Yes is chosen, the pump briefly displays a confirmation screen and returns
the user to the first screen of the Setup menu. Example screens are shown
here.
Note: The confirmation screen indicates whether Remote stop is enabled or
disabled, and displays the command sense of the remote control switch even if
Remote stop has been disabled
. This allows the command sense to be dis-
played if remote switching is disabled.

Even with the remote stop function disabled, the pump
could still start if the remote auto/manual toggle input
is used to toggle the pump into analogue mode.