Watson-Marlow 620U User Manual

Page 50

Advertising
background image

Watson-Marlow 620UN, 620U, 620SN, 620S User Manual

50

In the manual mode main screen, set the appropriate pump speed and direc-
tion using the UP or DOWN keys and the DIRECTION key. An example is
shown here. A slower speed may make it easier for the user to measure accu-
rately - though it may not represent the duty condition.
Place a suitable measuring vessel under the pump delivery tube.
Press START. The pump runs and fluid is pumped into the vessel.
When the required volume of fluid has been dispensed, it is necessary to stop
the pump and display the MemoDose screen. This can be achieved in several
ways.

1 Press the STOP key twice within half a second. The pump stops and
immediately displays the MemoDose screen.

OR... 2 Press the STOP key once. The pump stops. (This may make it easier to

ensure that the quantity of fluid that has been dispensed is sufficiently precise.)
Then press the STOP key twice within half a second. The pump displays the
MemoDose screen.

OR... 3 620UN, 620U Press the STOP key. The pump stops. Press the MENU

key. Use the UP or DOWN keys to select MemoDose. Press ENTER to con-
firm. The pump displays the MemoDose screen.

OR... 4 , 620SN, 620S Press the MemoDose key

17 MemoDose and calibration

Note: This menu area also gives access to flow rate calibration.

Each time the pump is started by pressing START, it records the number of pump-
head revolutions which occur until STOP is pressed. The number of revolutions is
proportional to the volume of fluid which has been dispensed: the dose. The
MemoDose facility allows the user to repeat-dose a precise volume of fluid. To do so,
a quantity of fluid must be dispensed as the master dose which the MemoDose facil-
ity can repeat exactly or proportionately.

To dispense a master dose

UN, U, SN, S

Advertising
This manual is related to the following products: