Xylem photoLab 6100 VIS User Manual
Page 122
Operation
photoLab
®
6100 VIS
122
ba75847e01
08/2009
Activating AQA1
The AQA1 monitoring is activated in the Mode menu:
Defining the AQA1
Interval
The AQA1 Interval defines the interval between two AQA1 checks.
When an interval has expired, the following consequences become effective:
z Warning and loss of the AQA1 labeling
z Locking of the photometer against all measurements (if activated).
Select and confirm Weeks.
AQA1 is active.
The Interval setting indicates
Weeks as the interval unit.
<HOME>
[AQA]
– AQA1 setup
– Mode
AQA1 setup
04/16/07 9:52
Mode
Weeks
Interval
Lock instrument
Configuration of tests ...
Mode
AQA1 inactive
Weeks
1
Enter a numeric value (2 to 52
weeks) (<0...9>) and confirm
The Interval defined for the AQA1
check is active.
<HOME>
[AQA]
– AQA1 setup
– Interval
AQA1 setup
04/16/07 9:52
Mode
Weeks
Interval
Lock instrument
Configuration of tests ...
Interval
2
Weeks