Xylem photoLab 6100 VIS User Manual

Page 132

Advertising
background image

Operation

photoLab

®

6100 VIS

132

ba75847e01

08/2009

Carrying out the

MatrixCheck

Note
The following description shows the proceeding for the MatrixCheck by
spiking. To switch to the MatrixCheck by dilution, use the [Dilute] function
key. The proceeding is similar there, but the entry of the Standard ID and
Standard concentration is not applicable.

1

Enter and confirm a numerical
value.

The setting is active.

2

Exit the menu with <ESC>.

AQA3/MatrixCheck setup

04/16/07 9:52

Maximum difference

10%

Maximum difference

10.0

%

1

Measure the original sample
without spiking or diluting it (see
section 4.5.1 to 4.5.3).

2

The measured value is displayed.

3

Open the setting menu with
[Setup].

4

Select and confirm AQA.

5

If necessary, check the settings in
the menu, AQA3/MatrixCheck
setup
.

6

Select and confirm AQA3/
MatrixCheck
.

The display for the MatrixCheck
opens up.

If the spiking with the standard
values of the CombiCheck R-2
suggested by the photometer
would cause the measuring range
to be exceeded, the MatrixCheck
by diluting is automatically
suggested.

Concentration

04/16/07 9:52

45

mg/l

18: C3/25

COD

16 mm

10 - 150 mg/l

Setup

Method list

Citation form

Unit

MatrixCheck (Spike)

04/16/07 9:52

Method

1: C3/25

Sample concentration

45 mg/lCOD

Standard ID

0

Standard concentration

0 mg/lCOD

Sample

Standard

Target value

[ml]

[ml]

[mg/l]

10

0

45

10

0

45

10

0

45

Dilute

Delete

Next