photoLab

®

6100 VIS

Operation

133

ba75847e01

08/2009

7

In theStandard ID entry field,
select the simplified MatrixCheck
with the CombiCheck standard
solution or enter a designation for
another standard solution used.

If the CombiCheck is selected, no
more entries are required
(continue with step 10).

8

Enter the concentration of the
used standard solution in the
Standard concentration entry field.

Specifying the series of
measurements:

9

Enter the volumes of sample and
standard of the individual test
sample solutions in the columns,
Sample [ml] and Standard [ml].
The nominal value is calculated
after each entry.

- You can delete a measurement

from the series of
measurements with [Delete].

Note that all nominal values have
to be within the measuring range
of the test.

10 Using [Next], accept all entries on

the page and switch to the next
page. The entries are checked by
the photometer.

The photometer is ready to carry
out the series of measurements.

Carrying out the series of
measurements:

According to the program, the
samples are measured top down.
You can, however, select the
samples yourself and thus change
the order with <

><>.

11 Use [Measurement] to proceed to

the measurement of the (first)
sample.

MatrixCheck (Spike)

04/16/07 9:52

Method

1: C3/25

Sample concentration

45 mg/lCOD

Standard ID

COD 1500

Standard concentration

400 mg/lCOD

Sample

Standard

Target value

[ml]

[ml]

[mg/l]

10

0.5

62

10

1

77

10

1.5

91

Dilute

Delete

Next

MatrixCheck (Spike)

04/16/07 9:52

Method

1: C3/25

Sample concentration

45 mg/lCOD

Sample

Standard

Target value

nominal

[ml]

[ml]

[mg/l]

[mg/l]

10

0.5

62

58

10

1

77

10

1.5

91

Back

Measureme

Complete