Maintenance and cleaning

151

ba75847e01

08/2009

5

Maintenance and cleaning

5.1

Replacing the lamp

Service life of the

tungsten halogen

lamp

The tungsten halogen lamp is a wear part with a certain average service life
(see chapter 7 T

ECHNICAL

DATA

). It has to be replaced if defective. The

photometer has a service hour counter for the lamp module (see section
4.18
).

Note
The replacement lamp is readily assembled as a lamp module and optically
adjusted in the factory. Therefore, treat it with utmost care. Fingerprints on
the glass will shorten the service life of the lamp. Do not touch the bulb of the
new lamp module with bare fingers. If you have touched the bulb
inadvertently, carefully clean it with a clean cloth soaked in alcohol.

Replacing the lamp

module

The lamp module is behind a cover (aluminium sheet) on the rear panel of
the photometer. Proceed as follows:

Caution

The lamp becomes very hot during operation. Do not touch the hot lamp
because it can cause burns! The lamp should cool down for approx.
10 minutes before it is exchanged.

1

Switch off the photometer and
disconnect it from the power
supply.

2

Unscrew the two screws (1) and
remove the lamp cover (2).

1

2