What to do if ...

157

ba75847e01

08/2009

Obviously incorrect

measured values

Fluctuating

measured values

Self test failed.

Connected printer

does not print

Cause

Remedy

– Cell contaminated

– Clean the cell

– Dilution set incorrectly

– Set the dilution

– Selected method not suitable

– Select different method

– Zero measurement incorrect

– Perform zero measurement

– Blank value incorrect

– Remeasure the blank value

– Cells that are not recognized (e.g.

some plastic cells) disturb the
AutoCheck measurement and, as
a consequence, falsify measured
values until the next AutoCheck is
carried out

– Use suitable cells (see section

7.1 and section 8.1)

Cause

Remedy

– Cell shaft cover open

– Close the cell shaft cover

Cause

Remedy

System test: Instrument defective

– Return instrument to service

department

Filter test: Instrument defective

– Return instrument to service

department

Lamp test: Lamp defective

– Replace lamp (see section 5.1)

– Return instrument to service

department

Wavelength calibration:

– Foreign particle in the cell shaft

– Instrument defective

– Clean the cell shaft

– Return instrument to service

department

Cause

Remedy

– Printer not suitable

– Connect a printer that can

interpret the printer control
language PCL-3