3 recording the kinetics, Recording the kinetics, Inner turn-up lid – Xylem photoLab 6100 VIS User Manual

Page 92

Advertising
background image

Operation

photoLab

®

6100 VIS

92

ba75847e01

08/2009

4.9.3

Recording the Kinetics

Note
During the recording, the photometer cannot carry out any regular self-test or
self-calibration (AutoCheck), because the recording would have to be
interrupted for this. A warm-up time of at least two hours is required for the
photometer to measure reliably during the recording.

Note the warm-up time of at least
2 hours for kinetic recordings.

1

Select the required profile with
[Profile] (see section 4.9.2).

The photometer is ready to
measure after the profile has been
selected.

2

Depending on the type, insert the
cell as follows:

Round cell:

Insert the round cell in the round
cell shaft so it touches the bottom.

If the inner turn-up lid is opened
too wide, a message prompts you
to close the inner turn-up lid.

<HOME>
Kinetics

Kinetics

04/16/07 9:52

Please select a profile for measuring!

Setup

Profile

Open

Kinetics

04/16/07 9:52

To start measurement,

insert cell or press <Start/Enter>

4002

Absorbance

Setup

Profile

Open

inner
turn-up lid