Operation photolab – Xylem photoLab 6100 VIS User Manual
Page 56

Operation
photoLab
®
6100 VIS
56
ba75847e01
08/2009
Measuring the
reagent blank value
Inserting a cell with barcode starts
a measurement.
If a cell without barcode is used:
Select the method manually
(see section 4.5.5).
The photometer is ready to
measure.
1
Using <ZERO·BLANK>, open the
Adjust selection list.
2
Select and confirm Blank value.
The window for the measurement
of the reagent blank value pops
up.
The data of the last measurement
appears in the measured value
display.
<HOME>
Concentration
Concentration
04/16/07 9:52
Please select method for measuring
or insert a barcoded cell
or insert AutoSelector.
Setup
Method list
Last method
New Method
Concentration
04/16/07 9:52
To start measurement,
insert cell or press <Start/Enter>
3: A6/25
NH
4
-N
16 mm
0.20 - 8.00 mg/l
Setup
Method list
Citation form
Unit
Concentration
04/16/07 9:52
3: A6/25
NH
4
-N
16 mm
0.20 - 8.00 mg/l
Setup
Method list
Citation form
Unit
Adjust
Zero adjustment
Blank value