5 measuring in concentration mode, 1 measuring cell tests with barcode, Measuring in – Xylem photoLab 6100 VIS User Manual

Page 41: Measuring cell tests with barcode

Advertising
background image

photoLab

®

6100 VIS

Operation

41

ba75847e01

08/2009

4.5

Measuring in Concentration mode

4.5.1

Measuring cell tests with barcode

Display if the value is not within the measuring range (see section 4.5.4).

Inserting a cell with barcode starts
a measurement.

1

Open the cell shaft cover.

2

Close the inner turn-up lid.

If the inner turn-up lid is opened
too wide, a message prompts you
to close the inner turn-up lid.

3

Insert the barcoded round cell in
the round cell shaft so it touches
the bottom. When doing so, align
the line mark with the notch at the
front of the round cell shaft.

The photometer selects the
method based on the bar code
and automatically starts
measurement.

4

Further options:

– Select a different citation form

with [Citation form],
(e.g. NH

4

<–> NH

4

-N).

– Select a different measuring

unit with [Unit],
(e.g. mg/l <–> mmol/l).

– Make further settings such as

dilution or blank value
measurements with [Setup]
(see section 4.5.6).

<HOME>
Concentration

Concentration

04/16/07 9:52

Please select method for measuring
or insert a barcoded cell
or insert AutoSelector.

Setup

Method list

Last method

New Method

Line mark

Barcode

Concentration

04/16/07 9:52

49

mg/l

18: C3/25

COD

16 mm

10 - 150 mg/l

Setup

Method list

Citation form

Unit