Xylem photoLab 6100 VIS User Manual
Page 46
Advertising
Operation
photoLab
®
6100 VIS
46
ba75847e01
08/2009
4.5.4
Exceeding the upper or lower limits of the measuring range
Measured value display if the measured value is outside the measuring
range:
MR = Measuring range
UL = Upper limit value of the measuring range
LL = Lower limit value of the measuring range
MV = Measured value
Range
Display
Example:
MR: 10 - 150 mg/l
LL < MV < UL
Measured value
1
UL < MV < UL + 10%
Upper limit of measuring
range exceeded by up to
10% and measured value
LL - 50% < MV < UL
Lower limit of measuring
range undercut by up to 50%
and measured value
2
MV > UL + 10%
Upper limit of measuring
range exceeded by more
than 10%
MV < LL - 50%
Lower limit of measuring
range undercut by more
than 50%
3
Invalid measured value
e.g. MV < 0