Xylem photoLab 6100 VIS User Manual
Page 55
photoLab
®
6100 VIS
Operation
55
ba75847e01
08/2009
User-defined
methods
For user-defined methods, you can activate the reagent blank value function
as follows only:
Note
If no value for E0 is stored during the entry of value pairs or the measurement
and storing of standard solutions for a nonlinear function (parabola or
polygon line), the message, No blank value correction is intended for this
method. appears when the User-defined blank value function is activated.
The blank value (E0) can be entered later by editing the method.
Entry type
Function type
Reagent
blank value
possible?
Entry of a function
(with and without entering the
ordinate intercept)
Linear
Yes
Nonlinear
No
Entry of value pairs or
measurement and storage of
standard solutions
(with entering or measuring and
storing E0)
Linear
Yes
Parabola
(second-order function)
Yes
Polygon line
No
Entry of value pairs or
measurement and storage of
standard solutions
(without entering or measuring
and storing E0)
Linear
Yes
Parabola
(second-order function)
Polygon line
Polygon line through
zero
No