Xylem photoLab 6100 VIS User Manual

Page 128

Advertising
background image

Operation

photoLab

®

6100 VIS

128

ba75847e01

08/2009

Activating AQA2

monitoring for a

method

Defining the AQA2

Interval, nominal

value and tolerance

The AQA2 Interval defines the interval between two AQA2 checks.
When an interval has expired, the following consequences become effective:

z Warning and loss of the AQA2 labeling

z Locking of the method against measurement (if activated).

Setting range:

1 to 12 weeks (default: 12 weeks) or

1 to 10000 measurements (default: 200 measurements)

Note
The unit of the AQA2 interval (Weeks or Measurements) is defined in the line,
Mode (see page 127).

Repeat the steps 1 to 8 if you want to configure further tests for AQA2.

1

Select a method (see section
4.5.3)
.

2

Select and confirm AQA2.

3

Select and confirm AQA2 active.

AQA2 is active for this method.

<HOME>
[AQA]

AQA2 setup

Method

...

AQA2 setup

04/16/07 9:52

Method

3: A6/25

AQA2

AQA2 active

Interval

12 Weeks

Target value

4.00 mg/l NH

4

-N

Tolerance

0.50 mg/l NH

4

-N

Standard ID

Method list

4

Select the Interval and enter the
AQA2 interval.

5

If necessary, adjust the values for
Target value and Tolerance.

6

Optional: Select Standard ID and
enter a designation. The
designation is recorded in the
AQA2 documentation.

AQA2 setup

04/16/07 9:52

Method

3: A6/25

AQA2

AQA2 active

Interval

12 Weeks

Target value

4.00 mg/l NH

4

-N

Tolerance

0.50 mg/l NH

4

-N

Standard ID

Method list