Technical data

159

ba75847e01

08/2009

7

Technical data

7.1

Measurement characteristics

Measuring principle

Single-beam spectrophotometer

Light source

Lamp type

Tungsten halogen lamp

Durability

approx. 1000 h

Monochromator

Type

Grating monochromator with step motor

Wavelength range

320 - 1100 nm

Max. scan speed

approx. 3300 nm/min

Wavelengths calibration

Automatic

Accuracy

± 1 nm

Reproducibility

± 0.5 nm
(checkable, e.g. with holmium oxide filter)

Resolution

1 nm

Spectral band width

4 nm

Photometric

measurement

Light sensor

Photo diode

Measuring range

A = -3.300 to A = +3.300

Linearity

< 1 % for A

≤ 2.000

in the range 340 to 900 nm

Accuracy *

– 0.003 A for A < 0.600

– 0.5 % of the reading for 0.600

≤A≤ 2.000

Reproducibility *

±

0.002 at A = 1.000

Resolution

Δ

A = 0.001

Scattered light

< 0.1 % transmission at 340 and 408 nm

* in the range 330 nm ... 1100 nm