7 technical data, 1 measurement characteristics, Technical data – Xylem photoLab 6100 VIS User Manual
Page 159: Measurement characteristics, Chapter 7 t, Echnical, Data, En in chapter 7 t, Nical, Ee section

photoLab
®
6100 VIS
Technical data
159
ba75847e01
08/2009
7
Technical data
7.1
Measurement characteristics
Measuring principle
Single-beam spectrophotometer
Light source
Lamp type
Tungsten halogen lamp
Durability
approx. 1000 h
Monochromator
Type
Grating monochromator with step motor
Wavelength range
320 - 1100 nm
Max. scan speed
approx. 3300 nm/min
Wavelengths calibration
Automatic
Accuracy
± 1 nm
Reproducibility
± 0.5 nm
(checkable, e.g. with holmium oxide filter)
Resolution
1 nm
Spectral band width
4 nm
Photometric
measurement
Light sensor
Photo diode
Measuring range
A = -3.300 to A = +3.300
Linearity
< 1 % for A
≤ 2.000
in the range 340 to 900 nm
Accuracy *
– 0.003 A for A < 0.600
– 0.5 % of the reading for 0.600
≤A≤ 2.000
Reproducibility *
±
0.002 at A = 1.000
Resolution
Δ
A = 0.001
Scattered light
< 0.1 % transmission at 340 and 408 nm
* in the range 330 nm ... 1100 nm