Xylem photoLab 6100 VIS User Manual

Page 152

Advertising
background image

Maintenance and cleaning

photoLab

®

6100 VIS

152

ba75847e01

08/2009

Note
After recommissioning, carry out a new zero adjustment for all
measurements.

3

Disconnect the electrical plug
connection (3). Disconnecting it
might take a lot of effort.

4

Unscrew the two knurled-head
screws (4) and remove the
defective lamp module (5).

5

Fix the new lamp module with the
knurled-head screws. When doing
so, the metal-plated side of the
PCB must point outward, i. e.
toward the knurled-head screws.

6

Reconnect the electrical plug
connection.

7

Fix the lamp cover.

8

Reset the service hour counter for
the lamp module (see section
4.16
).

3

4

5