Insert the cell with the blank
sample.
The first single measurement for
the reagent blank value takes
place.
The following data is displayed as
the result:
– The measured absorbance
from the (last) single
measurement.
– The median from all single
measurements carried out up to
now.
4
If necessary, carry out further
single measurements for the
formation of the median with [Next
meas.]
or
discard the last single
measurement with [Discard].
5
To accept the median value, press
[Apply].
The Lot number entry field pops
up.
6
Enter and confirm the Lot number
(<A...9>).
The blank value measurement is
completed.
The photometer is ready to
measure.
The use of the reagent blank value
is indicated by [BV/Lot number] in
the top right corner of the display.