The entered value for the dilution factor is valid for the selected method only.
The dilution factor is erased if
z the photometer is switched off
z a different method is selected
z the factor 0 is entered in the Dilution menu.
If a dilution factor is active, it is indicated on the display during measurement
in the form [1 + x].
1
Open the setting menu with
[Setup].
2
Select and confirm Dilution.
The input field for the dilution pops
up.
3
Enter and confirm the dilution
(<0...9>).
The entered dilution is taken into
account with the next
measurement.
Concentration
04/16/07 9:52
3: A6/25
NH
4
-N
16 mm
0.20 - 8.00 mg/l
Setup
Method list
Citation form
Unit
Sample + distilled water
1 + _