3 offset accuracy, 4 sine distortion – B&K Precision 4079 - Manual User Manual

Page 100

Advertising
background image

94

- Select OUT ON

8.

CHECK that the measured voltages on the DVM at 10Vp-p, 5Vp-p, 3Vp-p, 1Vp-p, 100 mVp-p and

50mVp-p are in the accuracy range:

Amplitude

setting

Minimum reading RMS Maximum reading RMS DVM reading RMS

10Vp-p

3.493V

3.578V

5Vp-p

1.743V

1.792V

3Vp-p

1.043V

1.078V

1Vp-p

0.343V

0.364V

100mVp-p

31mV

40mV

50mVp-p

14mV

22mV


5.3.3 Offset Accuracy

Specification: ±1% of programmed value ±10mV, into 50 ohm
Procedure:
1. Connect the OUTPUT connector to the input of the DVM using a 50 ohm coaxial cable and the

50 ohm 0.1% termination.

2. Set the DVM to measure DC volts.
3. Set the unit to generate a SINE wave signal with 1 KHz frequency. Set the unit amplitude to

10mVp-p

4. Set the unit offset to +4V.

7

CHECK that the measured voltages on the DVM at offsets of +4Vp-p, +2Vp-p, 0Vp-p, -2Vp-p,

-4Vp-p are in the accuracy range:

Offset setting

Minimum reading

Maximum reading

DVM reading

+4V

3.950V

4.050V

+2V

1.970V

2.030V

0V

-10mV

+10mV

-2V

-1.970V

-2.030V

-4V

-3.950V

-4.050V

5.3.4 Sine Distortion

Specification: - 65 dB to 20 KHz.

Procedure:

7.13 Connect the OUTPUT connector to the distortion analyzer using a 50 ohm coaxial cable

and a 50 ohm feedthrough termination.