2 mode key – B&K Precision 4079 - Manual User Manual

Page 23

Advertising
background image

17

F4:UNITS -

Selects the amplitude units: peak-to-peak, RMS or dBm.

F5:50 OHM/HI-Z

- Selects the amplitude voltage value based on the two different impedance termination. (i.e.

if connected to oscilloscope with 1Mohm, generator will display the correct amplitude value

for 1Mohm termination if HI-Z is selected)

F3:OFST - Selects the Offset parameter. Change the offset by using the cursor keys, rotary dial or numerical keypad.

If a certain setting cannot be produced, the waveform generator will display a “Setting Conflict” message.

F5:INTREF/EXTREF

- Selects internal or external reference source (the external reference must be connected to the

rear panel Ref In connector).

3.6.2 MODE Key

Selects the output mode: CONT (Continuous), TRIG (Triggered), GATE (Gated), and BRST (Burst).

To select the output mode, press MODE, then press the function key that corresponds to the desired Mode menu option, as

shown:

Mode Menu

F1: CONT - (Continuous) - Selects continuous output.

F2: TRIG - (Triggered) - Triggers one output cycle of the selected waveform for each trigger event.

F3: GATE - (Gated) - Triggers output cycles as long as the trigger source asserts the gate signal.

Front Panel

Amplitude

Setting

Data Point

Value

Relative Output

Amplitude Voltage

5 Vp-p

8191

+2.5 V

5 Vp-p

4095

+1.25 V

5 Vp-p

0

0V (offset voltage)

9 Vp-p

-4095

-4.5 V

4 Vp-p

-8191

-2 V