Considerations:

1) The value of the data at the start and end points must first be set by the user, using the

:ARB:DATA command. For example, suppose a line needs to be drawn from address 1000 to 4000 with data

point values 4000 and 8191 respectively like below:

To do this,

1. Set address to 1000 by sending ARB:ADDR 1000

2. Set data to 4000 by sending ARB:DATA 4000

3. Set address to 4000 by sending ARB:ADDR 4000

4. Set data to 8191 by sending ARB:DATA 8191

5. Then, use ARB:DRAW 1000,4000 to get the above line.

2) The range of the straight line cannot overlap with protected memory.

3) The end address must be greater than the start address.

4.13.4.5 Clear

:ARBitrary:CLEar <start address>,<end address>

This command is used to clear all or a portion of waveform memory. The memory is then set to the value zero.

Arguments

Type:

Numeric.

Numeric Range: 1 to 4,000,000

Rounding: to integer value

Command Type: Setting only.

Setting

Syntax:

:ARBitrary:CLEar<ws><start address>, <end address>

Examples: :ARB:CLE 1,1000

Considerations:

1) The clear range cannot overlap with protected memory.

2) The end address must be greater than the start address.

4.13.4.6 Copy

:ARBitrary:COPY <start address>,<length>,<destination address>

This command is used to copy a section of the waveform to a different location in waveform memory.

Arguments

Type:

NRf

Range:

1 to 4,000,000

Rounding:

to integer value

Command Type: Setting only

Setting

ADDR 1000

ADDR 4000

DATA 4000

DATA 8191

Line to Draw