1 pulse period, 2 pulse width, 3 pulse edge – B&K Precision 4079 - Manual User Manual

Page 69: 4 pulse rise

Advertising
background image

63

4.13.1.12.1 PULSe PERiod

This command sets the pulse period to the specified value.

NOTE: This also sets the pulse frequency.

Arguments

Type:

Numeric

Units:

S, mS, uS, nS

Range: 40nS-2000S

Rounding:

4 digits

Command Type: Setting or Query

Setting

Syntax:

[:SOURce:] PULse: PERiod <ws><value>

[:SOURce:] PULse: PERiod <ws>MINimum|MAXimum

Examples:

[:SOURce:] PULse: PERiod 500NS

Query

Syntax:

[:SOURce:] PULse: PERiod?[<ws>MINimum|MAXimum]

Response:

NR3

4.13.1.12.2 PULse WIDth

This command pulse width to the specified value.

Arguments

Type:

Numeric

Units:

S, mS, uS, nS

Range:

20 nS minimum; maximum defined by period and transition (see note above)

Rounding:

4 digits

Command Type: Setting or Query

Setting

Syntax:

[:SOURce:]PULse:WIDth<ws>< value >

[:SOURce:] PULse:WIDth <ws>MINimum|MAXimum

Examples:

[:SOURce:] PULse:WIDth 500NS

Query

Syntax:

[:SOURce:] PULse: WIDth?[<ws>MINimum|MAXimum]

Response:

NR3

4.13.1.12.3 PULse EDGe

This command sets both rising and falling edge of the pulse to the specified value.

Arguments

Type:

Numeric

Units:

S, mS, uS, nS

Range:

100 nS minimum; maximum defined by period and width (see note above)

Rounding:

4 digits

Command Type: Setting or Query

Setting

Syntax:

[:SOURce:] PULse:EDGe <ws>< value >

[:SOURce:] PULse:EDGe <ws>MINimum|MAXimum

Examples:

[:SOURce:] PULse:EDGe 500NS

Query

Syntax:

[:SOURce:] PULse:EDGe?[<ws>MINimum|MAXimum]

Response:

NR3

4.13.1.12.4 PULse RISe

This command sets rising edge of the pulse to the specified value.

Arguments

Type:

Numeric

Units:

S, mS, uS, nS

Range:

100 nS minimum; maximum defined by period and width (see note above)

Rounding:

4 digits