8 sync output - marker output - reference output – B&K Precision 4079 - Manual User Manual

Page 102

Advertising
background image

96

2. Push the MODE key on the unit front panel.
3. Verify that the oscilloscope displays a 1KHz sine wave. Make sure that the Output is ON.
4. Set the external function generator for a square wave output from 0V to 2V at 200 Hz.
5. Connect the function generator output to the unit TRIG IN connector.
6. Set the waveform generator:

Step 1. Set the output mode to triggered

MODE

F2 - TRIG

Step 2. Set the trigger source to internal

F2 - INT


7.

CHECK-Verify that the oscilloscope displays one cycle of sine wave per trigger event.

8.

Set the waveform generator:

Set the output mode to external trigger

F3 - PREV

F3 - EXT

9.

CHECK-Verify that the oscilloscope displays one sine waves per external trigger

10. Set the waveform generator:

Set the output mode to burst.

MODE

F4 - BRST

F2 - INT

11.

CHECK-Verify that the oscilloscope displays two cycles of sine wave per trigger event.

12. If you have a Model 4079 unit, connect the CH2 OUTPUT connector to the oscilloscope, push

CHAN key and repeat steps 2 through 13 for channel 2 (using CH2 control settings and CH2

TRIG IN connector).


5.3.8 Sync Output - Marker Output - Reference Output

Specification: TTL-level pulses at programmed frequency.
Procedure:

2) Connect the SYNC OUT connector to the oscilloscope vertical input.
3) Set the unit to generate sine waves at 1 KHz and

CHECK for a square waveform with

levels of 0V and >4V. on the oscilloscope.

4) Connect the REF OUT connector to the oscilloscope vertical input.
5)

CHECK for a 10 MHz square wave with 0V and >4V levels on the oscilloscope.

6) Connect the MARKER OUT connector to the oscilloscope vertical input.
7) Program the unit to generate an Arbitrary waveform with Marker:

- Press the WAVE key
- Select F5 – ARB
- Select F4 – EDIT
- Select F3 – PREDEF

Advertising