10 start address – B&K Precision 4079 - Manual User Manual

Page 78

Advertising
background image

72

Options:

SINusoid

SQUare

TRIangle

NOISe (Pseudo-Random Noise)

ANOise (Noise added to the current waveform)

Start Address

Type:

Numeric. The MIN and MAX forms both set the address to 1

Range:

1 to 4,000,000

Rounding:

to integer value

Length

Type:

Numeric.

Range

SIN:

16 to 65,536; divisible by 4

SQU:

2 to 65,536; divisible by 2

TRI:

16 to 65,536; divisible by 4

NOIS:

16 to 65,536

ANO:

16 to 65,536

Rounding:

to integer value.

Scale

Type:

Numeric. MIN sets the scale to 1; MAX sets the scale to 100

Range:

1 to 100 (See considerations)

Rounding;

to integer value.

Command Type: Setting only

Setting

Syntax:

:ARBitrary:PREDefined<ws> <shape>, <start>, <length>,<scale>

Examples:

:ARB:PRED SIN,1,1e3,100

Considerations:

1) The start address and the length must meet the specification that. Start address + Length - 1 ≤ 131,072

2) The 'scale' refers to the scaling of the waveform as a percentage of full scale. A scale of 100% will, under the correct

conditions, generate a waveform whose data values range from -8191 to +8191. These 'correct conditions' are set by the

'offset' value. This offset is the value of the data at the start address, and determines the maximum scale settable. The

following table shows the data values required in order to achieve maximum scale.

SHAPE

DATA

SIN

0

SQU

0

TRI

0

NOIS

0

4.13.4.10 Start Address

:ARBitrary:STARt <start address>

This command sets the start address of the waveform to be run.

Arguments

Type:

Numeric

Range:

1 to 3,999,999

Rounding: to integer value

Command Type: Setting or Query

Setting

Syntax:

:ARBitrary:STARt<ws><start address>

:ARBitrary:STARt<ws>MINimum|MAXimum