3 negative transition filter, 4 event register, 5 event enable register – B&K Precision 4079 - Manual User Manual

Page 83

Advertising
background image

77

Range: 0 to 131,072. Non integer arguments are rounded before execution.

Command Type: Setting or Query

Setting

Syntax:

:STAT:QUES:PTR<ws><NRf>

Examples: :STAT:QUES:PTR 2048

Query

Syntax:

:STAT:QUES:PTR?

Response:

NR1

4.13.5.4.3 Negative Transition Filter

:STAT:QUES:NTR

This command is used to set and query the value of the negative transition filter.

Arguments

Type:

NRf

Range:

0 to 131,072. Non integer arguments are rounded before execution

Command Type: Setting or Query

Setting

Syntax:

:STAT:QUES:NTR<ws><NRf>

Examples: :STAT:QUES:NTR 2048

Query

Syntax:

:STAT:QUES:NTR?

Response: NR1

4.13.5.4.4 Event Register

:STAT:QUES:EVENt?

This query is used to read the event register. Reading the register clears it.

Command Type: Query only

Query

Syntax:

:STATus:QUES:EVEN?

Response: NR1

4.13.5.4.5 Event Enable Register

:STAT:QUES:ENABle

This command is used to set and query the value of the enable register.

Arguments

Type:

NRf

Range:

0 to 131,072. Non integer arguments are rounded before execution.

Command Type: Setting or Query

Setting

Syntax:

:STAT:QUES:ENAB<ws><NRf>

Examples: :STAT:QUES:ENAB 2048

Query

Syntax:

:STAT:QUES:ENAB?

Response: NR1