1 output state, 2 output summing, 3 trigger subsystem – B&K Precision 4079 - Manual User Manual

Page 71: 1 trigger mode

Advertising
background image

65

Note: For model 4079, nothing changes in the commands above to control channel 1. But for channel 2, change :OUTP

to :OUTP2. For example, to change channel 2 output to ON, send the command: OUTP2 ON or OUTP2 1.

4.13.2.1 Output State

:OUTPut [:STATe] <state 0,1>

This command controls whether the output is ON or OFF.

Arguments

Type:

Boolean

Command Type: Setting or Query

Setting

Syntax:

:OUTPut[:STATe]<ws>ON|1|OFF|0

Examples:

:OUTP:STAT ON

:OUTP OFF

Query

Syntax:

:OUTPut[:STATe]?

Response:

0|1

4.13.2.2 Output Summing

:OUTPut :SUMming <state 0,1>

This command controls whether the summing is ON or OFF.

Arguments

Type:

Boolean

Command Type: Setting or Query

Setting

Syntax: :

OUTPut:SUMming<ws>ON|1|OFF|0

Examples:

:OUTP:SUM ON

:OUTP:SUM OFF

Query

Syntax:

:OUTPut:SUMming?

Response:

0|1

4.13.3 Trigger Subsystem

The Trigger Subsystem is used to control the waveform triggering. The command structure is as follows:

:TRIGger

:MODE CONTinuous|TRIGger|GATE|BURSt

:BURSt <numeric value>

:SOURce <MANual|INTernal|EXTernal|BUS

:TIMer <numeric value>

Note: For model 4079, nothing changes in the commands above to control channel 1. But for channel 2, change :TRIG

to :TRIG2. For example, to change channel 2 mode to gate, send the command: TRIG2:MODE GATE.

4.13.3.1 Trigger Mode

:TRIGger:MODE <trigger mode>

Advertising