B&K Precision 4079 - Manual User Manual

Page 104

Advertising
background image

98

Performance Tests Results

Setting

Minimum reading

Maximum reading

Reading

Frequency DDS

1 MHz

999990 Hz

1000010 Hz

Frequency ARB

1 MHz

999950 Hz

1000050 Hz

Sine Amplitude

10Vp-p

3.493V

3.578V

5Vp-p

1.743V

1.792V

3Vp-p

1.043V

1.078V

1Vp-p

0.343V

0.364V

100mVp-p

31mV

40mV

50mVp-p

14mV

22mV

Square Amplitude

10Vp-p

4.940V

5.060V

5Vp-p

2.465V

2.535V

3Vp-p

1.475V

1.525V

1Vp-p

0.485V

0.515V

100mVp-p

44mV

56mV

50mVp-p

20mV

31mV

ARB Amplitude

10Vp-p

3.493V

3.578V

5Vp-p

1.743V

1.792V

3Vp-p

1.043V

1.078V

1Vp-p

0.343V

0.364V

100mVp-p

31mV

40mV

50mVp-p

14mV

22mV

Offset Accuracy

+4V

3.950V

4.050V

+2V

1.970V

2.030V

0V

-10mV

+10mV

-2V

-1.970V

-2.030V

-4V

-3.950V

-4.050V

Sine Distortion

1 KHz

- 65 dB

Square Transition

1 MHz

6 ns

Variable Duty Cycle

20%

19.50%

20.50%

50%

49.50%

50.50%

80%

79.50%

80.50%

Advertising