1 frequency – B&K Precision 4079 - Manual User Manual
Page 59
53
:SHAPe SINusoid|SQUare|TRIangle
:FREQuency <numeric value>
:SOURce INTernal |EXTernal
:FSK
[:STATe] <Boolean>
:LOWFrequency <numeric value>
:HIFrequency <numeric value>
:RATE <numeric value>
:SOURce INTernal |EXTernal
:SWEep
:STATe <Boolean>
:SPACing <LIN|LOG>
:TIME <numeric value>
:STARt <numeric value>
:STOP <numeric value>
:PHAse
[:ADjust] <numeric value >
:PULSe
:PERiod <numeric value >
:WIDth <numeric value >
:EDGe <numeric value >
:RISe <numeric value >
:FALl <numeric value >
Note: For model 4079, nothing changes in the commands above to control channel 1. But for channel 2, change
:SOURce to :SOUR2. For example, to change channel 2 to SINE function, send the command: SOUR2:FUNC
SIN.
4.13.1.1 Frequency
:SOURce:FREQuency <frequency>
The frequency command controls the frequency of the output waveform.
Arguments
Type: Numeric.
Units: MHz, KHz, Hz (default)
Range: Dependent on the Point Rate and Wavelength.
Fmax = 1/(8nS * Wavelength)
Fmin = 1/(100S * Wavelength)
Rounding: The value is rounded to 4 digits.
Command Type: Setting or Query
Setting
Syntax: [:SOURce]:FREQuency[:CW]<ws><frequency>[units]
[:SOURce]:FREQuency<ws>MINimum|MAXimum
Examples: :FREQ 5KHZ
:FREQ 5E3
:FREQ MAXIMUM
:FREQ MIN
Query
Syntax: [:SOURce]:FREQuency[:CW]?[<ws>MAXimum|MINimum]
Examples: :FREQ?
:FREQ? MAX
Response: NR3
Considerations:
1) The MIN | MAX arguments should be used only in a Program Message that does NOT contain Program Message