2 device state, 1 local state (locs), 2 local with lockout state (lwls) – B&K Precision 4079 - Manual User Manual

Page 44: 3 remote state (rems), 4 remote with lockout state (rwls), 3 interface function subsets

Advertising
background image

38

User can return to local control with the press of any front panel keys, but it is extremely important to note that this

should be done ONLY when nothing is being sent or transferred between the instrument and the connected PC. Any

interruptions during transfer may delay the communication process, and in some cases freeze the instrument. For large

waveforms transmission, be it sending or receiving, please allow at least 15 seconds before pressing any front panel keys

to return to local mode. The instrument needs the additional time to completely generate or send the waveform.

The instrument accepts a carriage return (CR) as an end of string (EOS) terminator and sends both a CR and LF as the EOS

terminator.

4.2 Device State

The device may be in one of the four possible states described below. The transition between states is defined by IEEE 488.1.

4.2.1 Local State (LOCS)

In the LOCS the device may be operated from the front panel only. Its settings may be queried over the GPIB, but not

changed. Commands that do not affect the signal being output by the instrument are accepted.

4.2.2 Local With Lockout State (LWLS)

In the LWLS the device may be operated from the front panel only. Its settings may be queried over the GPIB, but not

changed. Commands that do not affect the signal being output by the instrument are accepted. The difference between the

LOCS and the LWLS is that from the LWLS the device may enter the Remote With Lockout State.

4.2.3 Remote State (REMS)

In the REMS the device may be operated from the GPIB. Actuating any front panel key will cause the device state to revert

to the LOCS.

4.2.4 Remote With Lockout State (RWLS)

In the RWLS the device is operable only from the GPIB. Front panel operation may be returned by either sending an

appropriate IEEE 488.1 command, or by cycling the device power.

4.3 Interface Function Subsets

The following interface function subsets are implemented in the MODEL 4076 and 4079: