10 sweep control, 1 sweep state, 2 sweep spacing – B&K Precision 4079 - Manual User Manual

Page 67: 3 sweep time, 4 sweep start

Advertising
background image

61

4.13.1.10 Sweep control

The following commands control the sweep functionality:

4.13.1.10.1 Sweep STATe

This command activates or deactivates sweep:

Arguments

Type:

Boolean

Command Type: Setting or Query

Setting

Syntax:

[:SOURce:]SWEEP[:STATe]<ws>ON|1|OFF|0

Examples:

SWEEP:STAT ON

SWEEP OFF

Query

Syntax:

[:SOURce:]SWEEP[:STATe]?

Response:

0|1

4.13.1.10.2 Sweep SPACing

This command sets the sweep spacing as either LINear or LOGarithmic:

Arguments

Type:

Character

Options:

LINear, LOGarithmic

Command Type: Setting or Query

Setting

Syntax:

[:SOURce:] SWEEP:SPACing<ws><LIN|LOG>

Examples:

[:SOURce:] SWEEP:SPACing LIN

SWEEP:SPAC LOG

Query

Syntax:

[:SOURce:] SWEEP:SPACing ?

Response:

LIN|LOG

4.13.1.10.3 Sweep TIME

This command sets the time for one complete sweep:

Arguments

Type:

Numeric

Units:

S, mS, uS, nS

Range:

10mS to 500S

Rounding:

to 4 digits

Command Type: Setting or Query

Setting

Syntax:

[:SOURce:]SWEEP:TIME<ws><time>[units]

[:SOURce:]SWEEP:TIME<ws>MINimum|MAXimum

Examples:

SWEEP:TIME 50MS

Query

Syntax:

[:SOURce:]SWEEP:TIME?[<ws>MINimum|MAXimum]

Response:

NR3

4.13.1.10.4 Sweep STARt

This command sets the start frequency of the sweep:

Arguments

Type:

Numeric.

Units:

MHz, KHz, Hz (default)

Range:

Dependent on the frequency range of the current function.

Rounding:

The value is rounded to 4 digits.