B&K Precision 4079 - Manual User Manual

Page 99

Advertising
background image

93

Amplitude

setting

Minimum reading RMS Maximum reading RMS DVM reading RMS

10Vp-p

3.493V

3.578V

5Vp-p

1.743V

1.792V

3Vp-p

1.043V

1.078V

1Vp-p

0.343V

0.364V

100mVp-p

31mV

40mV

50mVp-p

14mV

22mV


5. Set the unit to generate a SQUARE wave with 1KHz frequency.

6

. CHECK that the measured voltages on the DVM at 10Vp-p, 5Vp-p, 3Vp-p, 1Vp-p, 100 mVp-p and

50mVp-p are in the accuracy range:

Amplitude

setting

Minimum reading RMS Maximum reading RMS DVM reading RMS

10Vp-p

4.940V

5.060V

5Vp-p

2.465V

2.535V

3Vp-p

1.475V

1.525V

1Vp-p

0.485V

0.515V

100mVp-p

44mV

56mV

50mVp-p

20mV

31mV

4 Set the unit to the ARBITRARY function and generate a SINE wave signal:

- Press the WAVE key
- Select F5 – ARB
- Select F4 – EDIT
- Select F3 – PREDEF
- Select F2 – FROM, press key 1 and ENTER
- Select F2 – DATA, press key 0 and ENTER
- Select F3 – LENG, press 1000 and ENTER
- Select F4 – SCALE, press 100 and ENTER
- Select F5 – EXEC
- Select F3 – YES
- Select PARAM
- Select F1 – FREQ, press 1000 Hz