7.14 Set the unit to generate a sine waveform with 1KHz frequency and 10Vp-p.
7.15

Set the distortion analyzer to measure distortion in dB and select RMS response.

7.16

CHECK for a reading of more than -65 dB.

5.3.5 Square Transition Times

Specification: < 6 ns (10% to 90%) at 10Vp-p into 50 ohms.

Procedure:

1. Connect the OUTPUT connector to the oscilloscope input using a 50 ohm coaxial cable and a

50 ohm feedthrough termination.

2. Select WAVE and SQUARE to generate square waveform with 1MHz frequency and 10Vp-p.
3. Use the oscilloscope markers to measure the Rise and Fall times of the square, from 10% to

90%

levels.

4.

CHECK for rise and fall times of < 6 ns.



5.3.6 Square Variable Duty Cycle

Specification: variable from 20% to 80% with ± 0.5% accuracy at 1 KHz.

Procedure:

1. Connect the OUTPUT connector to the counter input using a 50 ohm coaxial cable and a 50 ohm

feedthrough termination.

2. Select WAVE and SQUARE to generate square waveform with 1KHz frequency and 5Vp-p.

3. Set the counter to measure duty cycle and select the automatic trigger mode.
4. On the unit change the square duty cycle to 20%, 50% and 80%
5.

CHECK for duty cycle readings between 19.50% - 20.50%, 49.50% - 50.50% and 79.50% - 80.50%.


5.3.7 Operating Modes

Specification: Continuous, triggered, gated, and burst.
Procedure:
1. Connect the OUTPUT connector to the oscilloscope vertical input using a 50 ohm coaxial

cable and a 50 ohm feed-through termination.