78

4.13.6 System Subsystem

The SYSTem subsystem collects the functions that are not related to instrument performance. The functions implemented in

the AWG are security, GPIB address changing, error queue reading, SCPI version reading, and power-on buffer setting (not

SCPI-defined). The command structure is as follows:

:SYSTem

:COMMunicate

:GPIB

:ADDRess <numeric value>

:ERRor?

:VERSion?

:SECurity

[:STATe] <Boolean>

:POBuffer <numeric value>

4.13.6.1 GPIB Address Change

:SYSTem:COMMunicate:GPIB:ADDRess

This command is used to set the GPIB address.

Arguments

Type:

Numeric

Range:

0 to 31

Rounding: to integer value

Command Type: Setting or Query

Setting

Syntax:

:SYSTem:COMMunicate:GPIB:ADDRess<ws><address>|MINimum|MAXimum

Example:

:SYST:COMM:GPIB:ADDR 20

Query

Syntax:

:SYSTem:COMMunicate:GPIB:ADDRess?[<ws>MINimum|MAXimum]

Response:

<address> in NR1 format

Considerations: 1) Setting the address to 31 puts the instrument in the off-bus state.

2) Using the MAX option sets the address to 30, not 31.

Default Power-on is address 9.

4.13.6.2 Error Queue Reading

:SYSTem:ERRor?

This query returns the first entry in the error queue, and removes that entry from the queue. Its function is identical to that of

the :STATus:QUEue:NEXT? query.

Command Type: Query only

Query

Syntax:

:SYSTem:ERRor?

Response:

<Error number>, "<error description>"

4.13.6.3 SCPI Version

:SYSTem:VERSion?