62

Command Type: Setting or Query

Setting

Syntax:

[:SOURce:]SWEEP:STARt<ws><frequency>[units]

[:SOURce:]SWEEP:STARt<ws>MINimum|MAXimum

Examples:

SWEEP:STARt 5KHZ

SWEEP:STARt 5E3

SWEEP:STARt MAXIMUM

SWEEP:STARt MIN

Query

Syntax:

[:SOURce:]SWEEP:STARt?[<ws>MAXimum|MINimum]

Examples:

SWEEP:STARt ?

SWEEP:STARt ? MAX

Response: NR3

4.13.1.10.5 Sweep STOP

This command sets the stop frequency of the sweep:

Arguments

Type:

Numeric.

Units:

MHz, KHz, Hz (default)

Range:

Dependent on the frequency range of the current function.

Rounding:

The value is rounded to 4 digits.

Command Type: Setting or Query

Setting

Syntax:

[:SOURce:]SWEEP:STOP<ws><frequency>[units]

[:SOURce:]SWEEP:STOP<ws>MINimum|MAXimum

Examples: SWEEP:STOP 5KHZ

SWEEP:STOP 5E3

SWEEP:STOP MAXIMUM

SWEEP:STOP MIN

Query

Syntax:

[:SOURce:]SWEEP:STOP?[<ws>MAXimum|MINimum]

Examples:

SWEEP:STOP ?

SWEEP:STOP ? MAX

Response: NR3

4.13.1.11 PHASe adjust

This command controls the phase adjustment

Arguments

Type:

Numeric

Units:

one (degrees implied)

Range:

-180 to +180 (other values converted to this range)

Rounding:

to integer

Command Type: Setting or Query

Setting

Syntax:

[:SOURce:]PHASe<ws><phase>

[:SOURce:]PHASe<ws>MINimum|MAXimum

Examples:

[:SOURce:]PHASe 500

Query

Syntax:

[:SOURce:]PHASe?[<ws>MINimum|MAXimum]

Response: NR3

4.13.1.12 PULSe setting

The following commands control the pulse function:

Note that width +0.6*(rise + fall) < period in order to have valid values. Width < Period – 10 ns