Luminex IS Version 2.3 User Manual

Page 100

Advertising
background image

Luminex IS Software Manual for Version 2.3

x

MAP Technology

90

PN 89-00002-00-110 Rev. B

2. Click OK. The system processes the various self- diagnostic

tests. When tests are complete, the Status Bar changes from a
“Processing” command state to an “Idle” command state. The
self-diagnostic tests should take less than one minute to
complete.

If the self diagnosis fails, you can obtain detailed information
regarding the results of the self-diagnostic test. See the following
“View Self-Diagnostic Details” section.

‹

To view details of the self-diagnostics test that passed or failed:

1. Click on the Diagnostics tab and view the Message log. If a

diagnostics test failed, an error message displays with a yellow
background.

2. Double-click the yellow row to see a detailed description. An

Errors dialog box opens showing a list of passed and failed self-
diagnostic tests. Click OK to close this dialog box.

Using the Cleanup

Utility

Use the Cleanup Utility to:

Perform a disk cleanup

Delete the Message Log Directory

Delete the Batch Directory

‹

To display the Cleanup Utility dialog box:

On the Tools menu, click Cleanup. The Cleanup Utility dialog
box opens. See Figure 62.

Figure 62 Cleanup Utility Dialog Box

‹

To perform a disk cleanup:

1. In the Cleanup Utility dialog box, click Disk Cleanup.

2. In the Select Drive dialog box, select the desired drive and click

OK. The Disk Cleanup dialog box opens showing it is
calculating progress. This can take several minutes.

3. When Windows finishes calculating the cleanup it displays the

Disk Cleanup for dialog box. See Figure 63. Check or uncheck

Advertising