Luminex IS Version 2.3 User Manual

Page 56

Advertising
background image

Luminex IS Software Manual for Version 2.3

x

MAP Technology

46

PN 89-00002-00-110 Rev. B

Figure 26 Update CON Targets Dialog Box

4. Ensure that the analyzer is set to draw CON1 and CON2 beads

from the wells you loaded in step 2 of “Run System xMAP
Calibrators” on page 43.

5. In the Maintenance tab, click CON1 then click OK. Wait until

CON1 completes. The Device Status section in the status bar
changes from “Running” to “Standby”.

6. Click CON2, then click OK. Wait until CON2 completes. The

Device Status section in the status bar changes from “Running”
to “Standby”. The System Monitor on the Diagnostics tab
displays date and time in green if both CON1 and CON2 are
successful.

7. Ensure the analyzer is set to the draw distilled water from the

well you loaded it in step 2 of “Run System xMAP Calibrators”
on page 43.

8. Click Wash to wash the system after running the system

calibrators. Wash a total of four times. You will need to change
the well location. Click on the dropdown menu located to the
right of the Wash button.

Advertising