MAP Technology

Luminex 2.3 Software

PN 89-00002-00-110 Rev. B

67

Managing Assay Lots

You can edit standard and control lot information. Once a lot is used,
changing or modifying it will prompt you for a new lot name. This
includes batches that have been set up, but not yet acquired.

You can modify known lot concentration values. If you change the
known concentration for a used lot, the system prompts you to enter
a new lot name. If you change the concentration values in an unused
lot, the system updates the lot with the new concentration values.

For assay reagents specified in templates, you can create new lots,
edit lot information, select pre-existing lots for reuse, import lots,
and export lots.

When editing lot numbers, follow these lot handling rules:

• If you have entered lot information for a template, but you

have not used the template to set up a batch, you can edit the
lot value.

• If you have entered lot information and have used the template

to set up a batch (even if it has not been acquired), you must
rename the lot.

Create a New Lot

‹

To create a new lot:

1. On the Home tab, click New Lot. The Open Template dialog

box opens. See Figure 41.

Figure 41 Open Template Dialog Box

2. Highlight the template that will use the new lot information and

click Select. Then, continue with substep “a” or” b”.

a.) If this is a new template with no associated lot information, a

New Standard Lot or New Control Lot dialog box opens as
appropriate. Enter the standard and control numbers to
continue. Once you enter the numbers, the Update Lot
Information
dialog box opens. See Figure 42 and Figure 43.

Advertising