Luminex IS Software Manual for Version 2.3

x

MAP Technology

10

PN 89-00002-00-110 Rev. B

heater block temperature is out of range or experiencing
unacceptable temperature conditions, including these items:

Luminex XYP heater block temperature time-out

heat circuit failure

temperature out-of-range sensing

temperature change since calibration, or a change in channel
temperature

Table 1 defines the values listed in the System Monitor. These values
are useful for diagnostic purposes when communicating with
Luminex Technical Support.

Table 1. System Monitor Values

Property

Value Units of Measure

Air Pressure

PSI, air pressure to the sheath
container from the air pump

Sheath Pressure

PSI, sheath pressure from the
sheath container through the
system

Delta Cal Temperature

°C, temperature deviation from
the last calibration

Board Temperature

°C, temperature of the analog
board

DD Temperature

°C, DD temperature at the U
block inside the optics platform

CL1 Temperature

°C, CL1 temperature at the U
block inside the optics platform

CL2 Temperature

°C, CL2 temperature at the U
block inside the optics platform

XYP Board Temperature

°C, temperature of the XYP
board inside the Luminex XYP
instrument

XYP Heater Temperature

°C, temperature of the XYP
heaters inside the Luminex
XYP instrument

XYP Heater Temperature In
Range

Indicates if the XYP heater is in
the set range