Maintenance procedures – Luminex IS Version 2.3 User Manual

Page 98

Advertising
background image

Luminex IS Software Manual for Version 2.3

x

MAP Technology

88

PN 89-00002-00-110 Rev. B

information for CAL 1, CAL2, CON1, and CON2 reagents are
correct.

5. Click OK to verify that the lot information is accurate.

Maintenance Procedures

Table 11 shows a recommended schedule for maintenance
operations.

Table 11. Maintenance Operations:

Recommended Use Schedule

Operation

Recommended Use Schedule

Warmup

Daily

After four hours of system inactivity

Prime

Daily

To remove air from sheath fluid tubing

After performing these actions:

– refilling the sheath container

– removing and replacing sheath container

– changing the sheath fluid filter

– changing the syringe seal

Backflush

Troubleshooting and preventative
maintenance purposes only:

– to remove obstructions from the cuvette

– if fluid does not flow through the waste

tubing during prime cycles or during
sample acquisition

– if fluid drips from the sample probe during

priming and forms puddles of fluid on the
plate

Alcohol Flush

Daily

Before system calibration

After changing the sample probe

To remove air bubbles from the cuvette using
70% isopropanol or 70% ethanol

Sanitize

To decontaminate sample lines and cuvette
after biohazard contact using 10% to 20%
household bleach

– daily if working with biohazards

– monthly if not working with biohazards

Wash

As needed using distilled water

Four times after system calibration

Twice after sanitize