3 disagreement between indication and actual flow – Yokogawa ADMAG AXF User Manual

Page 101

Advertising
background image

IM 01E20C01-01E

10-5

10. MAINTENANCE

10.3.3 Disagreement Between Indication and Actual Flow

NO

NO

NO

NO

YES

NO

NO

YES

YES

YES

NO

YES

YES

NO

NO

YES

Set the parameters correctly.

Examine the condition of the fluid in the
flowtube, of bubbles, and of grounding.

Execute zero adjustment when the
flowtube is filled completely with fluid and
when the fluid is not moving.

• Change the mounting position or piping

so that the flowtube is
completely filled with fluid.

• Switch to vertical mounting.

Install a defoaming device on the
flowtube’s upstream side.

Completely ground the converter and
flowtube (Refer to the section on
grounding in the relevant user’s manual).

Use within the specified conductivity range.

Confirm that a valve is located
immediately upstream of the flowtube
(i.e., on the upstream side).

• Remove scales. Especially clean the

electrodes and their vincinity.

• Use a soft brush or cloth to perform

scale removal, and make sure that the
lining is not damaged.

YES

• If any valve located immediately

upstream of the flowtube is used in a
half-open condition, the indication may
be affected by turbulence in the flow.

• Relocate the valve to the downstream

side.

• If the valve must be located on the

upstream side, ensure that the minimum
distance to the flowmeter is 10 times
the flow-tube diameter; alternatively,
use the valve fully open.

START

Are parameters

set correctly?

Is the tube

filled completely with

fluid?

Is the

flowmeter completely

grounded?

Is the

fluid likely to contain

deposits?

Is a valve

located immediately upstream

of the flowtube?

Examine the operation of the converter
using an AM012 calibrator in order to
determine whether the fault lies in the
converter or in the flowtube.

Contact a service center.

Is the

signal cable

terminal treatment

complete?

Fully terminate the signal cable using
PVC tubing or insulation tape, thus
insuring that conductors, inner shields,
and outer shields do not make contact
with each other or with the case.

Signal terminals A-C and B-C:

100 M

⍀ or more?

Excitation terminals EX1-C:

1 M

⍀ or more?

NO

Dry the tube fully to restore insulation
resistance.

YES

Was zero

adjustment carried out

correctly?

Does the fluid

contain bubbles?

F1007.EPS

YES

Does the

conductivity stay within the

specified range?

Remove the flowtube and examine the
insulation resistance.

Advertising