Maintenance, 1 maintenance, 1 fuse replacement – Yokogawa ADMAG AXF User Manual
Page 97: 2 setting of switches, 1 lifting of display unit, Maintenance -1, 1 fuse replacement -1, Setting of switches -1, 1 lifting of display unit -1

IM 01E20C01-01E
10-1
10. MAINTENANCE
10. MAINTENANCE
10.2 Setting of Switches
IMPORTANT
• Lifting of display unit is necessary for setting
switches. Perform display unit lifting as de-
scribed in section 10.2.1.
• To preserve the safety, do not touch the
electrical circuit and the cables except the
setting switches.
10.2.1 Lifting of display unit
(1) Turn off the power.
(2) Remove the cover as described in section 4.4.1.
(3) Loosen the amplifier assembly’s two screws while
supporting it with your hand (See Figure 10.2.1).
(4) Lift the display unit (See Figure 10.2.2).
At this time, do not remove the connector.
(5) Set the switches as described in Section 10.2.2 and
10.2.3.
(6) Taking care not to entangle the cables, install the
display unit with two mounting screws.
(7) Install the cover.
F1008.EPS
Display unit mounting
screws (x2)
Figure 10.2.1 Removing the Front Cover and Mounting
Screws of Display Unit
10.1 Maintenance
WARNING
• Maintenance work must be carried out by the
trained personnel having knowledge of safety
standard and not by operators.
• When opening the cover, wait for more than 10
minutes after turning off the power. Further-
more, opening of the cover must also be
carried out by the trained personnel having
knowledge of safety standard.
IMPORTANT
• As a rule, maintenance of this flowmeter should
be implemented in a maintenance service shop
where the necessary tools are provided.
• The amplifier assembly contains sensitive parts
that may be damaged by static electricity.
Excercise care so as not to directly touch the
electronic parts or circuit patterns on the board,
for example, by preventing static electrification
by using grounded wrist straps when handing
the assembly.
10.1.1 Fuse Replacement
CAUTION
Please contact Yokogawa's service office for
fuse replacement. Also be sure to use the fuse
that was supplied by Yokogawa’s sales or
service offices.